ERBE Erbotom ICC 300 Features

The ERBE ICC 300 is a high-frequency surgical unit for cutting and coagulation. It's the price-conscious solution for medium and major surgical procedures. After all, a smooth routine in the operating room depends on maximum safety, a high degree of comfort in use, and perfect functionality - exactly the features of the ICC 300 . And it also meets the special demands of Minimally Invasive Surgery.






  • CUT CONTROL for automatically perfect cuts.Reproducible cut quality with minimum necrosis

  • HlGH CUT with specialized arc regulation for underwater cuts or in fatty structures

  • SOFT coagulation for precisely controlled coagulation

  • No carbonization (SOFT COAG)

  • Far less adhesion and electrode soiling

  • Bipolar coagulation for even greater safety

  • PPS for intelligent support of initial incisions and cutting

  • AUTO START for greater facility of bipolar coagulation

  • SPRAY coagulation for better surface effects

  • ARGONPLASIVIA coagulation via integrated plug-in connection

ERBE Erbotom ICC 300 Specifications

Dimensions

  • Dimesnsions (HxWxD) cm, (In) (Display) 41 x 15.2 x 36.8 (16.1 x 5.9 x 14.5)

  • Weights, kg (lb) (Display) 10 (22

Electrical

  • Monopolar (Systems) Cut(blend), coagulate

  • Max. energy output ( at rated ohms) (SYSTEMS) More than 250 Watt 300

  • Max voltage, vp-p (Systems) 1200

  • Max. energy output ( at rated ohms) (SYSTEMS) Up to 150 Watt 120

  • Max voltage, vp-p (Systems) 380 soft, 5,600 spray

  • Bipolar (SYSTEMS) Coagulate

  • Max. energy output ( at rated ohms) (SYSTEMS) Up to 150 Watt 120

  • Max voltage, vp-p (Systems) 380

Power Connection

  • Rated power voltage 240 V / 230 V / 115 V / 110 V / 100 V ± 10 %

  • Rated power frequency 50 / 60 Hz

  • Power current 4.0 A at 230 - 240 V / 8.0 A at 100 - 115 V

  • Power consumption in Standby mode 25 watts

  • Power consumption at max. HF power 620 watts 920 VA

  • Current consumption in Standby mode 150 mA at 230 - 240 V / 300 mA at 100 - 115 V

  • Potential equalization terminal yes

  • Power fuses 2 slow burn, 4 A at 230 - 240 V / 8 A at 100 - 115 V

Cutting, monopolar with automatic voltage control (Auto Cut)

  • HF voltage waveform unmodulated sinusoidal alternating voltage

  • Crest factor C, at RL = 500 ohms C = 1.4 for all setttings

  • Rated frequency 330 kHz

  • Maximum HF voltage at load resistance RL = ∞ 650 VP

  • Dynamic internal impedance in rated load range 0 ohm

  • Constancy of the HF voltage at RL> 350 ohms automatically controlled

  • Cutting quality 4 coagulation effects selectable by key

  • Constancy of the 4 coagulation effects automatically controlled

  • HF rated power 300 watts at RL = 500 ohms

  • HF power limitation 1 to 300 watts in 1 watt steps

  • Setting of the HF power limitation via up/down keys

  • Display of the HF power limitation 7-segment display, 3 decimal places

  • Precision of the HF power limitation +/- 1 digit or +/- 15%

  • Activation of Cutting mode by key or pedal

  • HF connecting sockets 2, electronically locked from one another

Cutting, monopoly with automatic arc control (High Cut)

  • HF voltage waveform unmodulated sinusoidal alternating voltage

  • Crest factor C, at RL = 500 ohms C = 1.4 for all setttings

  • Rated frequency of the HF voltage 330 kHz

  • Maximum HF voltage at load resistance RL = ∞ 570 VP

  • Constancy of the intensity of the electric arc automatically controlled

  • Cutting quality 4 coagulation effects selectable by key

  • Constancy of the 4 coagulation effects automatically controlled

  • HF rated power 300 watts at RL = 500 ohms

  • HF power limitation 1 to 300 watts in 1 watt steps Setting of the

  • HF power limitation via up/down keys Display of the

  • HF power limitation 7-segment display, 3 decimal places

  • Precision of the HF power limitation +/- 1 digit or +/- 15%

  • PPS (Power Peak System) yes

  • Activation of Cutting mode by key or pedal

  • HF connecting sockets 2, electronically locked from one another

Soft Coagulation (Soft Coag)

  • HF voltage waveform unmodulated sinusoidal alternating voltage

  • Crest factor C, at RL = 500 ohms C = 1.4 for all setttings

  • Rated frequency of the HF voltage 330 kHz

  • Peak value of the HF voltage max. 1900 VP

  • Constancy of the HF voltage automatically controlled

  • HF rated power 120 watts at 125 ohms

  • HF power limitation (PHFmax.) from 1 watt to 120 watts in 1 watt steps

  • Setting of the HF power limitation via up/down keys

  • Display of the HF power limitation 7-segment display, 3 decimal places

  • Precision of the HF power limitation +/- 1 digit or +/- 15%

  • Activation of Cutting mode by key or pedal

  • HF connecting sockets 2, electronically locked from one another

 
 
 